Implication of CT table sag on geometrical accuracy during virtual simulation.
Document Type
Academic Journal
Author
Zullo JR; Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston, TX 77030, USA. jrzullo@mdanderson.org; Kudchadker RWu RLee APrado K
Source
Publisher: Elsevier Science Country of Publication: United States NLM ID: 8908862 Publication Model: Print Cited Medium: Print ISSN: 0958-3947 (Print) Linking ISSN: 18734022 NLM ISO Abbreviation: Med Dosim Subsets: MEDLINE
Subject
Language
English
ISSN
0958-3947
Abstract
Computed tomography (CT) scanners are used in hospitals worldwide for radiation oncology treatment simulation. It is critical that the process very accurately represents the patient positioning to be used during the administration of radiation therapy to minimize the dose delivery to normal tissue. Unfortunately, this is not always the case. One problem is that some degree of vertical displacement, or sag, occurs when the table is extended from its base when under a clinical weight load, a problem resulting from mechanical limitations of the CT table. In an effort to determine the extent of the problem, we measured and compared the degree of table sag for various CT scanner tables at our institution. A clinically representative weight load was placed on each table, and the amount of table sag was measured for varying degrees of table extension from its base. Results indicated that the amount of table sag varied from approximately 0.7 to 6.6 mm and that the amount of table sag varied not only between tables from different manufacturers but also between tables of the same model from the same manufacturer. Failure to recognize and prevent this problem could lead to incorrectly derived isocenter localization and subsequent patient positioning errors. Treatment site-specific and scanner-based laser offset correction should be implemented for each patient's virtual simulation procedure. In addition, the amount of sag should be measured under a clinically representative weight load upon CT-simulator commissioning.