Clinical image quality in daily practice of breast cancer mammography screening.
Document Type
Academic Journal
Author
Guertin MH; Institut National de Santé Publique du Quebec, Québec, Québec, Canada. Electronic address: marie-helene.guertin@inspq.qc.ca.; Théberge I; Institut National de Santé Publique du Quebec, Québec, Québec, Canada.; Dufresne MP; Hôpital Maisonneuve-Rosemont, Montréal, Québec, Canada.; Zomahoun HT; Centre de Recherche du CHU de Québec, Axe Santé des populations et pratiques optimales en santé, Québec, Québec, Canada.; Major D; Institut National de Santé Publique du Quebec, Québec, Québec, Canada.; Tremblay R; Ministère de la Santé et des Services sociaux, Québec, Québec, Canada.; Ricard C; Clinique Radiologique Audet Inc, Québec, Québec, Canada.; Shumak R; Cancer Care Ontario, Toronto, Ontario, Canada.; Wadden N; Memorial University, Faculty of Medicine, St John's, Newfoundland, Canada.; Pelletier E; Institut National de Santé Publique du Quebec, Québec, Québec, Canada.; Brisson J; Centre de Recherche du CHU de Québec, Axe Oncologie, Québec, Québec, Canada; Département de Médecine Sociale et Préventive, Faculté de médicine and Centre de Recherche sur le cancer, Université Laval, Québec, Québec, Canada.
Source
Publisher: SAGE Publications Country of Publication: United States NLM ID: 8812910 Publication Model: Print-Electronic Cited Medium: Internet ISSN: 1488-2361 (Electronic) Linking ISSN: 08465371 NLM ISO Abbreviation: Can Assoc Radiol J Subsets: MEDLINE
Subject
Language
English
Abstract
Objective: To assess the quality of screening mammograms performed in daily practice in the Quebec Breast Cancer Screening Program.
Subjects and Methods: Clinical image quality of a random subsample of 197 screening mammograms performed in 2004-2005 was independently evaluated by 2 radiologists based on the criteria by Canadian Association of Radiologists (CAR). When disagreement occurred for overall judgement or positioning score, the mammograms were reviewed by a third radiologist. Cohen's kappas for interrater agreement were computed. Multivariable robust Poisson regression models were used to study associations of overall quality and positioning with body mass index (BMI) and breast density.
Results: The CAR criteria were not satisfied for 49.7% of the mammograms. Positioning was the quality attribute most often deficient, with 37.2% of mammograms failing positioning. Interrater agreement ranged from slight (kappa = 0.02 for compression and sharpness) to fair (kappa = 0.30 for exposure). For overall quality, women with a BMI ≥ 30 kg/m(2) had a failure proportion of 67.5% compared with 34.9% for women with a BMI<25 kg/m(2) (risk ratio 2.1 [95% confidence interval, 1.5-3.0]). For positioning, women with a BMI ≥ 30 kg/m(2) had a failure proportion of 53.8% compared with 27.9% for women with a BMI < 25 kg/m(2) (risk ratio 1.9 [95% confidence interval, 1.2-3.1]). Effects of breast density on image quality differed among radiologists.
Conclusion: Despite measures to ensure high-quality imaging, including CAR accreditation, approximately half of this random sample of screening mammograms failed the CAR quality standards. It would be important to define quality targets for screening mammograms carried out in daily practice to interpret such observations.
(Copyright © 2014 Canadian Association of Radiologists. Published by Elsevier Inc. All rights reserved.)