Field Evaluation of a TSI Model 3034 Scanning Mobility Particle Sizer in New York City: Winter 2004 Intensive Campaign.
Document Type
Article
Source
Aerosol Science & Technology. Oct2006, Vol. 40 Issue 10, p753-762. 10p. 2 Diagrams, 1 Chart, 8 Graphs.
Subject
*PARTICLE size determination
*PARTICLE counting (Water treatment plants)
*PARTICLE size distribution
*PARTICLES
*CONDENSATION
*SCANNING systems
*STATISTICAL correlation
Language
ISSN
0278-6826
Abstract
A new “single box” Scanning Mobility Particle Sizer (TSI SMPS Model 3034) was deployed and operated during a period of four weeks as a part of the PMTACS-NY Winter 2004 intensive study in Queens College, New York City. The SMPS 3034 is an alternative to a conventional multi-component TSI SMPS and houses a Differential Mobility Analyzer and butanol-based Condensation Particle Counter in one cabinet. The SMPS 3034 operates at a fixed 1 l/min sample flow rate (4 l/min sheath flow rate) and measures size distributions within a 10–487 nm size range. One size scan is produced every 3 minutes. Four other measurement systems (a conventional TSI SMPS with a Nano Differential Mobility Analyzer, an Aerodynamic Particle Sizer, a stand-alone Condensation Particle Counter, and an R&P Inc. Filter Dynamic Measurement System (FDMS) TEOM mass monitor) were operated side-by-side with the SMPS 3034. It is shown that total particle number concentrations measured by the SMPS 3034 are highly correlated with those from the conventional Nano SMPS, the Condensation Particle Counter and the FDMS TEOM monitor, and that the number median diameters measured by the SMPS 3034 and the Nano SMPS agree within 3 nm. [ABSTRACT FROM AUTHOR]