학술논문

A Single-Round Infection Fluorescent SARS-CoV-2 Neutralization Test for COVID-19 Serological Testing at a Biosafety Level-2 Laboratory.
Document Type
Article
Source
Viruses (1999-4915). Jun2022, Vol. 14 Issue 6, pN.PAG-N.PAG. 12p.
Subject
*NEUTRALIZATION tests
*SERODIAGNOSIS
*COVID-19 testing
*SARS-CoV-2
*COVID-19
*VIRAL antibodies
Language
ISSN
1999-4915
Abstract
A robust serological test to measure neutralizing antibodies against SARS-CoV-2 in biosafety level-2 (BSL-2) laboratories is useful for monitoring antibody response after vaccination or natural infection. The gold standard assay is the conventional plaque reduction neutralization test (PRNT) which requires extensive labor, live viruses, and BSL-3 facilities. Recently, we developed a novel single-round infection fluorescent SARS-CoV-2 virus (SFV) that can be safely used at BSL-2 laboratories for high-throughput neutralization and antiviral testing. In this study, we evaluated the performance of the neutralization test using this SFV with 80 PRNT-positive and 92 PRNT-negative clinical serum or plasma specimens. The SFV neutralization test (SFVNT) has 100% sensitivity and specificity compared to the PRNT. Furthermore, the neutralizing titers generated by the SFVNT and PRNT are highly correlated, with R2 = 0.903 (p < 0.0001). Due to high sensitivity, specificity, accuracy, and reproducibility, the SFVNT can be deployed for the large-scale testing of COVID-19 patients or vaccinated people in general lab settings. [ABSTRACT FROM AUTHOR]