Clinical and Technical Validation of Novel Bite Force Measuring Device for Functional Analysis after Mandibular Reconstruction.
Document Type
Article
Source
Diagnostics (2075-4418). Feb2023, Vol. 13 Issue 4, p586. 11p.
Subject
*FUNCTIONAL analysis
*FREE flaps
*PLASTIC surgery
*SURGICAL flaps
*CAPACITIVE sensors
*FIBULA
MANDIBLE surgery
Language
ISSN
2075-4418
Abstract
Bite force measuring devices that are generally suitable for edentulous patients or patients undergoing mandibular reconstruction are missing. This study assesses the validity of a new bite force measuring device (prototype of loadpad®, novel GmbH) and evaluates its feasibility in patients after segmental mandibular resection. Accuracy and reproducibility were analyzed with two different protocols using a universal testing machine (Z010 AllroundLine, Zwick/Roell, Ulm, Germany). Four groups were tested to evaluate the impact of silicone layers around the sensor: no silicone ("pure"), 2.0 mm soft silicone ("2-soft"), 7.0 mm soft silicone ("7-soft") and 2.0 mm hard silicone ("2-hard"). Thereafter, the device was tested in 10 patients prospectively who underwent mandibular reconstruction using a fibula free flap. Average relative deviations of the measured force in relation to the applied load reached 0.77% ("7-soft") to 5.28% ("2-hard"). Repeated measurements in "2-soft" revealed a mean relative deviation of 2.5% until an applied load of 600 N. Maximum bite force decreased postoperatively by 51.8% to a maximum mean bite force of 131.5 N. The novel device guarantees a high accuracy and degree of reproducibility. Furthermore, it offers new opportunities to quantify perioperative oral function after reconstructive surgery of the mandible also in edentulous patients. [ABSTRACT FROM AUTHOR]