학술논문

Image acquisition and interpretation criteria for Tc-HMPAO-labelled white blood cell scintigraphy: results of a multicentre study.
Document Type
Article
Source
European Journal of Nuclear Medicine & Molecular Imaging. Apr2014, Vol. 41 Issue 4, p615-623. 9p. 1 Color Photograph, 2 Diagrams, 6 Charts.
Subject
*RADIONUCLIDE imaging
*RADIOLABELED leucocytes
*PHYSIOLOGICAL effects of isotopes
*INFECTION
*NUCLEAR medicine
*DIAGNOSIS
Language
ISSN
1619-7070
Abstract
Purpose: There is no consensus yet on the best protocol for planar image acquisition and interpretation of radiolabelled white blood cell (WBC) scintigraphy. This may account for differences in reported diagnostic accuracy amongst different centres. Methods: This was a multicentre retrospective study analysing 235 WBC scans divided into two groups. The first group of scans (105 patients) were acquired with a fixed-time acquisition protocol and the second group (130 patients) were acquired with a decay time-corrected acquisition protocol. Planar images were interpreted both qualitatively and semiquantitatively. Three blinded readers analysed the images. Results: The most accurate imaging acquisition protocol comprised image acquisition at 3 - 4 h and at 20 - 24 h in time mode with acquisition times corrected for isotope decay. Conclusion: Using this protocol, visual analysis had high sensitivity and specificity in the diagnosis of infection. Semiquantitative analysis could be used in doubtful cases, with no cut-off for the percentage increase in radiolabelled WBC over time, as a criterion to define a positive scan. [ABSTRACT FROM AUTHOR]