Prospective Laboratory Evaluation of the cobas Plasma Separation Card for HIV and Treponema pallidum Antibody Analysis.
Document Type
Article
Source
Sexually Transmitted Diseases. Nov2023, Vol. 50 Issue 11, p764-769. 6p.
Subject
Language
ISSN
0148-5717
Abstract
Background: The cobas Plasma Separation Card (PSC; Roche Diagnostics) was developed for HIV viral load testing. This study evaluates the performance of HIVand Treponema pallidum (Tp) antibody (Ab) detection on PSCs as an alternative to dried blood spots (DBSs). Methods: EDTA whole blood samples were collected from HIV-positive (n = 100), HIV-negative (n = 50), Tp-positive (n = 100), and Tp-negative patients (n = 50) and spotted on DBS and PSC. Antibody detection performance was evaluated for HIVAb using the Genscreen ULTRA HIVAg-Ab test (Bio-Rad) and for Tp Ab using the Syphilis Total Ab test (Bio-Rad). Plasma was used as a reference specimen. Results: Receiver operating characteristic curve analysis for DBS and PSC generated areas under the curve (AUC + 95% confidence interval) of 0.985 (0.960--1.000) and 0.987 (0.973--1.000) for HIV Ab and 1.000 (1.000--1.000) and 0.996 (0.983--1.000) for Tp Ab, respectively. Receiver operating characteristic areas under the curve were not significantly different between DBS and PSC for HIV or TPAb. At selected cutoff values rendering at least 99% sensitivity for HIV Ab detection, the specificity was 96% on DBS and 68% on PSC. For Tp Ab detection at 90% sensitivity, 100% specificity is reached on both DBS and PSC (exceeding the required 95%). However, the median quantitative HIVand Tp Ab signal of positive samples significantly decreased in PSC compared with DBS and plasma. Conclusions: Although receiver operating characteristic analysis does not seem to indicate significant differences in performance between DBS and PSC, the significant reduction in quantitative Ab detection signal dictates card composition optimization before its use for HIV and TpAb detection can be advised. [ABSTRACT FROM AUTHOR]