Evaluation of partial pressure CO2 change in the dialyzer blood inlet during hemodialysis as a measure of vascular access recirculation.
Document Type
Article
Source
Hemodialysis International. Oct2023, Vol. 27 Issue 4, p370-377. 8p.
Subject
*ARTERIAL catheterization
*PARTIAL pressure
*HEMODIALYSIS
*HEMODIALYSIS patients
*SURVIVAL rate
Language
ISSN
1492-7535
Abstract
Introduction: Vascular access recirculation during hemodialysis is associated with reduced effectiveness and worse survival outcomes. To evaluate recirculation, an increase in pCO2 in the blood of the arterial line during hemodialysis (threshold of 4.5 mmHg) was proposed. The blood returning from the dialyzer in the venous line has significantly higher pCO2, so in the presence of recirculation, pCO2 in the arterial blood line may increase (ΔpCO2) during hemodialysis sessions. The aim of our study was to evaluate ΔpCO2 as a diagnostic tool for vascular access recirculation in chronic hemodialysis patients. Methods: We evaluated vascular access recirculation with ΔpCO2 and compared it with the results of a urea recirculation test, which is the gold standard. ΔpCO2 was obtained from the difference in pCO2 in the arterial line at baseline (pCO2T1) and after 5 min of hemodialysis (pCO2T2). ∆pCO2 = pCO2T2–pCO2T1. Findings: In 70 hemodialysis patients (mean age: 70.52 ± 13.97 years; hemodialysis vintage of 41.36 ± 34.54, KT/V 1.4 ± 0.3), ∆pCO2 was 4 ± 4 mmHg, and urea recirculation was 7% ± 9%. Vascular access recirculation was identified using both methods in 17 of 70 patients, who showed a ∆pCO2 of 10 ± 5 mmHg and urea recirculation of 20% ± 9%; time in months of hemodialysis was the only difference between vascular access recirculation and non‐vascular access recirculation patients (22 ± 19 vs. 46 ± 36, p: 0.05). In the non‐vascular access recirculation group, the average ΔpCO2 was 1.9 ± 2 (p: 0.001), and the urea recirculation % was 2.8 ± 3 (p: 0.001). The ΔpCO2 correlated with the urea recirculation % (R: 0.728; p < 0.001). Discussion: ΔpCO2 in the arterial blood line during hemodialysis is an effective and reliable diagnostic tool for identifying recirculation of the vascular access but not its magnitude. The ΔpCO2 test application is simple and economical and does not require special equipment. [ABSTRACT FROM AUTHOR]