Meeting the European Commission performance criteria for the use of triple quadrupole GC-MS/MS as a confirmatory method for PCDD/Fs and dl-PCBs in food and feed samples.
Document Type
Article
Source
Analytical & Bioanalytical Chemistry. May2016, Vol. 408 Issue 13, p3511-3525. 15p.
Subject
*FURANS
*DIOXINS
*TETRACHLORODIBENZODIOXIN
*CHROMATOGRAPHIC analysis
Language
ISSN
1618-2642
Abstract
Until recently, European Union (EU) legislation required the use of high-resolution gas chromatography coupled to high-resolution mass spectrometry (HRGC-HRMS) based on magnetic sector analyzers as a standard approach for confirmatory analysis of dioxins (PCDDs) and furans (PCDFs) in feed and food. However, recent technological advances in MS instruments enabled other alternative analytical techniques to meet the same analytical criteria as those requested for HRGC-HRMS. In this sense, triple quadrupoles (GC-MS/MS) can be a realistic alternative for the analysis of dioxins. In this work, the performance of GC-MS/MS technology was evaluated against the criteria demanded by the EU for confirmatory analysis of dioxins and PCBs in food and feed. Thus, the study comprises a number of parameters including chromatographic separation, limit of quantification, linearity, repeatability, and ion ratio precision. Analyses of solvent standards as well as sample extracts (inter-calibration extracts and certified reference materials) were also considered within the scope of this study. Additionally, direct comparisons of the results obtained by GC-MS/MS with those from GC-HRMS were made. The results of this work suggested that GC-MS/MS was highly sensitive and selective for confirmatory analysis of PCDD/Fs and related compounds in food and feed samples and meets all the criteria requested by the European Commission. [ABSTRACT FROM AUTHOR]