The timed 'Up & Go' is dependent on chair type.
Document Type
Article
Source
Clinical Rehabilitation. Sep2002, Vol. 16 Issue 6, p609-616. 8p. 1 Black and White Photograph, 3 Charts, 1 Graph.
Subject
*MOTOR ability
*ERGONOMICS
*MEDICAL rehabilitation
Language
ISSN
0269-2155
Abstract
Introduction: The timed 'Up & Go' (TUG) is a performance test identifying problems in functional mobility. More knowledge on how the type of chair used influences test results is needed. Objective: To investigate inter-rater agreement on the time score and to assess if chair type used influenced the performance of the test. Setting: (1) Inter-rater agreement investigation on the time score was carried out with elderly individuals living in a retirement home (n = 31). (2) Four types of chairs were tested on elderly individuals in three different health care centres (n = 100). Results: The two observers were close in timing (mean difference = 0.04 s). From a reference chair the median time for TUG was 15.7 s compared with 16.9 s from a chair with a low seat (p < 0.001). It was significantly more difficult to stand up from a chair without armrests (p < 0.001), and from the lowest chair (p < 0.001), which was also the only chair difficult to sit down on (p = 0.02). Conclusion: The inter-rater agreement of the time scoring of the TUG has been confirmed. Test performance is dependent on chair type; chairs with armrests and a seating height of 44-47 cm should be used. Clinicians must follow standard procedures and equipment when using the test or else risk invalidating test findings. [ABSTRACT FROM AUTHOR]