Optically sectioned fluorescence endomicroscopy with hybrid-illumination imaging through a flexible fiber bundle.
Document Type
Article
Source
Journal of Biomedical Optics. May2009, Vol. 14 Issue 3, p030502-030502-3. 1p.
Subject
*FLUORESCENCE microscopy
*OPTICAL resolution
*LIGHTING
*MEDICAL imaging systems
*OPTICAL instruments
*MEDICAL microscopy
*LABORATORY rats
Language
ISSN
1083-3668
Abstract
We present an endomicroscope apparatus that exhibits out-of-focus background rejection based on wide-field illumination through a flexible imaging fiber bundle. Our technique, called HiLo microscopy, involves acquiring two images, one with grid-pattern illumination and another with standard uniform illumination. An evaluation of the image contrast with grid-pattern illumination provides an optically sectioned image with low resolution. This is complemented with high-resolution information from the uniform illumination image, leading to a full-resolution image that is optically sectioned. HiLo endomicroscope movies are presented of fluorescently labeled rat colonic mucosa. [ABSTRACT FROM AUTHOR]