Reduction of electromagnetic interference from a commercially available MR-compatible flow simulator.
Document Type
Journal Article
Source
Journal of Medical Engineering & Technology. Jan/Mar2000, Vol. 24 Issue 1, p28-31. 4p.
Subject
*ELECTROMAGNETIC interference
*MAGNETIC resonance imaging
*MAGNETIC resonance imaging equipment
*COMPARATIVE studies
*ELECTROMAGNETISM
*RESEARCH methodology
*MEDICAL cooperation
*IMAGING phantoms
*RESEARCH
*EVALUATION research
Language
ISSN
0309-1902
Abstract
Electromagnetic interference (EMI) associated with the electronics of a commercially available computer controlled flow simulator substantially decreases the quality of the MR image. The effect of a custom-built radiofrequency shield on its spectral emission, and the corresponding signal-to-noise ratio measured for the image of a standard phantom, were determined. The results demonstrate the elimination of EMI and a significant improvement in image quality. [ABSTRACT FROM AUTHOR]