학술논문

Capture-stabilize approach for membrane protein SPR assays.
Document Type
Article
Source
Scientific Reports. 12/12/2014, p1-9. 9p.
Subject
*MEMBRANE proteins
*DYNAMICS
*SURFACE plasmon resonance
*MONOCLONAL antibodies
*CONFORMATIONAL analysis
Language
ISSN
2045-2322
Abstract
Measuring the binding kinetics of antibodies to intact membrane proteins by surface plasmon resonance has been challenging largely because of the inherent difficulties in capturing membrane proteins on chip surfaces while retaining their native conformation. Here we describe a method in which His-tagged CXCR5, a GPCR, was purified and captured on a Biacore chip surface via the affinity tag. The captured receptor protein was then stabilized on the chip surface by limited cross-linking. The resulting chip surface retained ligand binding activity and was used for monoclonal antibody kinetics assays by a standard Biacore kinetics assay method with a simple low pH regeneration step.Wedemonstrate the advantages of this whole receptor assay when compared to available peptide-based binding assays. We further extended the application of the capture-stabilize approach to virus-like particles and demonstrated its utility analyzing antibodies against CD52, a GPI-anchored protein, in its native membrane environment. The results are the first demonstration of chemically stabilized chip surfaces for membrane protein SPR assays. [ABSTRACT FROM AUTHOR]