Comparative analysis of syringes versus pump devices in benchtop aspiration thrombectomy models: A systematic review and meta-analysis.
Document Type
Article
Source
Interventional Neuroradiology. Dec2023, p1.
Subject
Language
ISSN
1591-0199
Abstract
Although direct contact aspiration has emerged as one of the leading techniques for mechanical thrombectomy (MT), there is still ongoing debate about the aspiration/suction pump devices that can optimize recanalization rates. To address this gap, we conducted a meta-analysis comparing the aspiration efficacy of 60 ml syringe and pump devices in benchtop MT models.Systematic literature review was conducted using Medline, Embase, Web of Science, and Scopus in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analysis guidelines. Outcomes of interest included flow rate and vacuum pressure delivered by a 60 ml syringe and several aspiration pumps. We used a random effects model to calculate the mean difference (MD) with 95% confidence intervals (CIs) and a statistically significant difference was considered as a two-sided p-value of less than 0.05.We included six benchtop studies comparing 60 ml syringes and vacuum pumps. Our meta-analysis showed that there were no significant differences in vacuum pressure (MD:0.71inHg, 95% CI: [−0.81;2.23], p = 0.359) and flow rate (MD:0.27 mL/s, 95% CI: [−3,07; 3.61], p = 0.873) between 60 ml syringes and vacuum pumps groups.Our study demonstrated comparable performance in terms of vacuum pressure and flow rates between a 60 ml syringe and a heterogeneous combination of commercially available aspiration pumps. [ABSTRACT FROM AUTHOR]