Optimization of the fluid-attenuated inversion recovery (FLAIR) imaging for use in autopsy imaging of the brain region using synthetic MRI.
Document Type
Article
Source
Technology & Health Care. 2023, Vol. 31 Issue 2, p661-674. 14p.
Subject
*BRAIN imaging
*MAGNETIC resonance imaging
*AUTOPSY
*CEREBROSPINAL fluid
*DEAD
Language
ISSN
0928-7329
Abstract
BACKGROUND: The failure of cerebrospinal fluid (CSF) signal suppression in postmortem fluid-attenuated inversion recovery (FLAIR) of the brain is a problem. OBJECTIVE: The present study was to clarify the relationship between the temperature of deceased persons and CSF T 1 , and to optimize the postmortem brain FLAIR imaging method using synthetic MRI. METHODS: Forehead temperature was measured in 15 deceased persons. Next, synthetic MRI of the brain was performed, the CSF T 1 was measured, and the optimal TI was calculated. Two types of FLAIR images were obtained with the clinical and optimal TI. The relationship between forehead temperature and the CSF T 1 and optimal TI was evaluated. The optimized FLAIR images were physically and visually evaluated. RESULTS: The CSF T 1 and optimal TI were strongly correlated with forehead temperature. Comparing the average SNR and CNR ratios and visual evaluation scores of the two FLAIR images, those captured with the optimal TI showed statistically lower SNR, higher CNR, and higher visual evaluation scores (p < 0.01). CONCLUSIONS: Synthetic MRI enables the quantification of the CSF T 1 resulting from postmortem temperature decreases and calculation of the optimal TI, which could aid in improving the failure of CSF signal suppression and in optimizing postmortem brain FLAIR imaging. [ABSTRACT FROM AUTHOR]