Development and field testing of a rapid and ultra-stable atmospheric carbon dioxide spectrometer.
Document Type
Article
Source
Atmospheric Measurement Techniques Discussions. 2014, Vol. 7 Issue 8, p8101-8123. 23p.
Subject
*SPECTROMETER design & construction
*ATMOSPHERIC carbon dioxide & the environment
*GAS cylinders
Language
ISSN
1867-8610
Abstract
We present field test results for a new spectroscopic instrument to measure atmospheric carbon dioxide (CO2) with high precision (0.02 ppm at 1 Hz) and demonstrate high stability (within 0.1 ppm over more than 8 months), without the need for hourly, daily, or even monthly calibration against high-pressure gas cylinders. The technical novelty of this instrument (ABsolute Carbon dioxide, ABC) is the spectral null method using an internal quartz reference cell with known CO2 column density. Compared to a previously described prototype, the field instrument has better stability and benefits from more precise thermal control of the optics and more accurate pressure measurements in the sample cell (at the mTorr level). The instrument has been deployed at a long-term ecological research site (the Harvard Forest, USA), where it has measured for eight months without on-site calibration and with minimal maintenance, showing drift bounds of less than 0.1 ppm. Field measurements agree well with those of another commercially available cavity ring-down CO2 instrument (Picarro G2301) run with a standard calibration protocol. This field test demonstrates that ABC is capable of performing high-accuracy, unattended, continuous field measurements with minimal use of calibration cylinders. [ABSTRACT FROM AUTHOR]