SPR and Double Resonance LPG Biosensors for Helicobacter pylori BabA Antigen Detection.
Document Type
Article
Source
Sensors (14248220). Apr2024, Vol. 24 Issue 7, p2118. 12p.
Subject
*HELICOBACTER pylori
*SURFACE plasmon resonance
*HELICOBACTER pylori infections
*RESONANCE
*BIOSENSORS
*ANTIGENS
*MEMBRANE proteins
Language
ISSN
1424-8220
Abstract
Given the medical and social significance of Helicobacter pylori infection, timely and reliable diagnosis of the disease is required. The traditional invasive and non-invasive conventional diagnostic techniques have several limitations. Recently, opportunities for new diagnostic methods have appeared based on the recent advance in the study of H. pylori outer membrane proteins and their identified receptors. In the present study we assess the way in which outer membrane protein–cell receptor reactions are applicable in establishing a reliable diagnosis. Herein, as well as in other previous studies of ours, we explore the reliability of the binding reaction between the best characterized H. pylori adhesin BabA and its receptor, the blood antigen Leb. For the purpose we developed surface plasmon resonance (SPR) and double resonance long period grating (DR LPG) biosensors based on the BabA–Leb binding reaction for diagnosing H. pylori infection. In SPR detection, the sensitivity was estimated at 3000 CFU/mL—a much higher sensitivity than that of the RUT test. The DR LPG biosensor proved to be superior in terms of accuracy and sensitivity—concentrations as low as 102 CFU/mL were detected. [ABSTRACT FROM AUTHOR]