Absorption Mode FTICR Mass Spectrometry Imaging.
Document Type
Article
Source
Analytical Chemistry. 12/3/2013, Vol. 85 Issue 23, p11180-11184. 5p.
Subject
*ION cyclotron resonance spectrometry
*MAGNETIC fields
*SIGNAL-to-noise ratio
*FOURIER transforms
Language
ISSN
0003-2700
Abstract
Fourier transform ion cyclotron resonance mass spectrometry offers the highest mass resolving power for molecular imaging experiments. This high mass resolving power ensures that closely spaced peaks at the same nominal mass are resolved for proper image generation. Typically higher magnetic fields are used to increase mass resolving power. However, a gain in mass resolving power can also be realized by phase correction of the data for absorption mode display. In addition to mass resolving power, absorption mode offers higher mass accuracy and signal-to-noise ratio over the conventional magnitude mode. Here, we present the first use of absorption mode for Fourier transform ion cyclotron resonance mass spectrometry imaging. The Autophaser algorithm is used to phase correct each spectrum (pixel) in the image, and then, these parameters are used by the Chameleon work-flow based data processing software to generate absorption mode "Datacubes" for image and spectral viewing. Absorption mode reveals new mass and spatial features that are not resolved in magnitude mode and results in improved selected ion image contrast. [ABSTRACT FROM AUTHOR]