학술논문

Improvements of a low-cost CO2 commercial NDIR sensor for UAV atmospheric mapping applications.
Document Type
Article
Source
Atmospheric Measurement Techniques Discussions. 4/19/2022, p1-20. 20p.
Subject
*MOLE fraction
*TRACE gases
*DRONE aircraft
*DETECTORS
*CLINICAL pathology
Language
ISSN
1867-8610
Abstract
Unmanned Aerial Vehicles (UAVs) provide a cost-effective way to fill in gaps between surface in-situ observations and remote-sensed data from space. In this study, a novel portable CO2 measuring system suitable for operations on-board small-sized UAVs has been developed and validated. It is based on a low-cost commercial nondispersive near-infrared (NDIR) 15 CO2 sensor (Senseair AB, Sweden), with a total weight of 1058 g, including batteries. The system performs in situ measurements autonomously, allowing for its integration into various platforms. Accuracy and linearity tests in the lab showed that the precision remains within ±1 ppm (1s) at 1 Hz. Corrections due to temperature and pressure changes were applied following environmental chamber experiments. The accuracy of the system in the field was validated against a reference instrument (Picarro, USA) onboard a piloted aircraft and it was found to be ±2 ppm (1s) at 1 Hz and ±1 ppm (1s) at 1 min. 20 Due to its fast response, the system has the capacity to measure CO2 mole fraction changes at 1 Hz, thus allowing the monitoring of CO2 emission plumes and the characteristic of their spatial and temporal distribution. Details of the measurement system and field implementations are described to support future UAV platform applications for atmospheric trace gas measurements. [ABSTRACT FROM AUTHOR]