LC-MS/MS method for the quantification of masitinib in RLMs matrix and rat urine: application to metabolic stability and excretion rate.
Document Type
Article
Source
Chemistry Central Journal. Dec2017, Vol. 11 Issue 1, p1-7. 7p.
Subject
*PROTEIN-tyrosine kinase inhibitors
*KINASE inhibitors
*LIQUID chromatography
*MASS spectrometry
*PHARMACOKINETICS
*IONS
Language
ISSN
1752-153X
Abstract
Masitinib (MST) is a selective tyrosine kinase inhibitor. Validated liquid chromatography tandem mass spectrometric method (LC-MS/MS) was developed for the quantification of MST in rat liver microsomes (RLMs) matrix. The developed method was applied to metabolic stability and excretion rate studies. Reversed phase liquid chromatography was used for resolution of MST and bosutinib (IS) using C (50 mm × 2.1 mm, 1.8 μm). Binary solvent system consisted of 35% solvent A (0.1% formic acid in HO, pH: 3.2) and 65% solvent B (acetonitrile) used as mobile phase at flow rate of 0.25 mL with a total run time of 5 min. Injection volume was 5 µL. Generation of ions was done in positive ESI source and quantification of MST and IS were done using MRM mode. The developed method showed a linearity in the range of 5-200 ng/mL (r ≥ 0.9992) with LOQ and LOD of 0.25 and 0.76 ng/mL in RLMs. The intra- and inter-day precision and accuracy ranged from 0.95 to 1.49 and − 5.22 to 1.13%, respectively in RLMs. Rate of disappearance of MST during incubation with RLMs was almost linear allover incubation time. In vitro t was 50.38 min and CL was 3.11 ± 0.2. The developed method was applied also to measure the rate of masitinib excretion in rat urine. The method can used for further pharmacokinetic studies of MST. [ABSTRACT FROM AUTHOR]