Hemodynamic Assessment of Hollow-Fiber Membrane Oxygenators Using Computational Fluid Dynamics in Heterogeneous Membrane Models.
Document Type
Article
Source
Journal of Biomechanical Engineering. May2021, Vol. 143 Issue 5, p1-14. 14p.
Subject
*COMPUTATIONAL fluid dynamics
*OXYGENATORS
*HEMODYNAMICS
*BLOOD flow
*EXTRACORPOREAL membrane oxygenation
*SHEARING force
Language
ISSN
0148-0731
Abstract
Extracorporeal membrane oxygenation (ECMO) has been used clinically for more than 40 years as a bridge to transplantation, with hollow-fiber membrane (HFM) oxygenators gaining in popularity due to their high gas transfer and low flow resistance. In spite of the technological advances in ECMO devices, the inevitable contact of the perfused blood with the polymer hollow-fiber gas-exchange membrane, and the subsequent thrombus formation, limits their clinical usage to only 2–4 weeks. In addition, the inhomogeneous flow in the device can further enhance thrombus formation and limit gas-transport efficiency. Endothelialization of the blood contacting surfaces of ECMO devices offers a potential solution to their inherent thrombogenicity. However, abnormal shear stresses and inhomogeneous blood flow might affect the function and activation status of the seeded endothelial cells (ECs). In this study, the blood flow through two HFM oxygenators, including the commercially available iLA® MiniLung Petite Novalung (Xenios AG, Germany) and an experimental one for the rat animal model, was modeled using computational fluid dynamics (CFD), with a view to assessing the magnitude and distribution of the wall shear stress (WSS) on the hollow fibers and flow fields in the oxygenators. This work demonstrated significant inhomogeneity in the flow dynamics of both oxygenators, with regions of high hollow-fiber WSS and regions of stagnant flow, implying a variable flow-induced stimulation on seeded ECs and possible EC activation and damage in a biohybrid oxygenator setting. [ABSTRACT FROM AUTHOR]