Evaluating Indirect Blood Pressure Measurement Techniques: A Comparison of Three Systems in Infants and Children.
Document Type
Article
Source
Pediatrics. Sep78, Vol. 62 Issue 3, p326. 5p.
Subject
*BLOOD pressure measurement
*PEDIATRICS
Language
ISSN
0031-4005
Abstract
ABSTRACT. The systemic arterial blood pressures obtained in infants and children utilizing three indirect measuring devices--Arteriosonde 1010, Infrasonde 3000, Pedisphyg system--were compared to each other and to intra-arterial pressure measured directly. The results indicate that Arteriosonde performs considerably better than Infrasonde; nevertheless, the Arteriosonde values are often only approximations of true systole and diastole. The Pedisphyg system yields accurate, reproducible values for systole; however, the system is not designed to determine diastolic blood pressure. [ABSTRACT FROM AUTHOR]