Predictive value of pulse oximetry and fetal scalp blood pH in the case of meconium-stained amniotic fluid
Document Type
Journal Article
Source
European Journal of Obstetrics & Gynecology & Reproductive Biology. Jul2003, Vol. 109 Issue 1, p27-32. 6p.
Subject
*CORD blood
*MECONIUM
*AMNIOTIC liquid
*COMPARATIVE studies
*FETUS
*HYDROGEN-ion concentration
*INFANT mortality
*LABOR (Obstetrics)
*RESEARCH methodology
*MECONIUM aspiration syndrome
*MEDICAL cooperation
*OXIMETRY
*OXYGEN
*RESEARCH
*SCALP
*EVALUATION research
*RECEIVER operating characteristic curves
Language
ISSN
0301-2115
Abstract
Objective: To compare the predictive value of intrapartum fetal pulse oximetry to that of fetal scalp blood pH for an abnormal neonatal outcome in the case of thick meconium.Study Design: A prospective multicenter observational study was performed from June 1994 to November 1995. Fetal oxygen saturation was monitored using a Nellcor N-400 fetal pulse oximeter in case of abnormal FHR. The last readings of fetal oxygen saturation and fetal scalp blood pH before birth were used to assess the ability of both techniques to predict an abnormal neonatal status.Results: At a 30% cutoff, the negative predictive value of fetal oxygen saturation was not altered in case of meconium when compared to clear amniotic fluid (79 and 83%, respectively). Fetal scalp blood pH at a 7.20 threshold had a poor negative predictive value in case of meconium when compared to clear amniotic fluid (56% versus 88%, respectively). The receiver operator characteristic curve showed similar performance of fetal scalp blood pH and pulse oximetry in cases with clear amniotic fluid. In cases with meconium, the performance of fetal scalp blood pH was poor, whereas that of pulse oximetry remained unchanged. In three cases with meconium below the vocal cords, a drop in fetal oxygen saturation was observed during labor whereas fetal scalp blood pH remained within normal values.Conclusion: The predictive value of fetal scalp blood pH is poor in case of meconium, whereas the predictive value of pulse oximetry seems to be unchanged in this situation. [ABSTRACT FROM AUTHOR]