Non-inductive plasma vertical position measurement for the 1056 s discharge on EAST.
Document Type
Article
Source
Review of Scientific Instruments. 10/1/2022, Vol. 93 Issue 10, p1-4. 4p.
Subject
*ELECTROMAGNETS
*TOKAMAKS
*POLARISCOPE
Language
ISSN
0034-6748
Abstract
Vertical position stability plays a crucial role in maintaining safe and reliable plasma operation for long-pulse fusion devices. In general, the vertical position is measured by using inductive magnetic coils installed inside the vacuum vessel; however, the integration drift effects are inherent for steady-state or long-pulse plasma operation. Developing a non-magnetic approach provides a fusion reactor-relevant steady-state solution that avoids the negative impact of integration drift. In this paper, we compare the non-inductively determined vertical position achieved by line-integrated interferometer and polarimeter measurements to that employing an inductive flux loop for a 1056 s discharge recently achieved on EAST (Experimental Advanced Superconducting Tokamak). Experimental results show that the non-inductive measurement is more robust than flux loops after 300 s if the integrator is not reset to suppress integrator drift. Real-time vertical position control using the non-inductive system is proposed for the next EAST experimental campaign. [ABSTRACT FROM AUTHOR]