Ex vivo hemolysis: Three cases demonstrating mechanically-induced hemolysis from the extracorporeal circuit during hemodialysis.
Document Type
Article
Source
Clinical Biochemistry. Jun2023, Vol. 116, p133-137. 5p.
Subject
*HEMOLYSIS & hemolysins
*ERYTHROCYTES
Language
ISSN
0009-9120
Abstract
Sporadic mechanically-induced hemolysis associated with kinks in extracorporeal blood circuits during hemodialysis is a rare but potentially serious complication that exhibits laboratory features consistent with both in vivo and in vitro hemolysis. Misclassification of clinically significant hemolysis as in vitro can lead to inappropriate test cancellation and delayed medical interventions. Here, we report three cases of hemolysis attributed to kinked hemodialysis blood lines, which we have defined as " ex vivo " hemolysis. All three cases demonstrated an initial mixed picture of laboratory features consistent with both classifications of hemolysis. Specifically, absent features of in vivo hemolysis on blood film smear despite normal potassium led to the misclassification of these samples as in vitro hemolysis and their cancellation. A proposed mechanism for these overlapping laboratory features is the recirculation of damaged red blood cells from the kinked or pinched hemodialysis line back into the patient circulation producing an " ex vivo " hemolysis presentation. In two of the three cases, the patients developed acute pancreatitis as a result of hemolysis and required urgent medical follow up. We developed a decision pathway to help laboratories in identifying and handling these samples by recognizing that in vitro and in vivo hemolysis have overlapping laboratory features. These cases highlight the need for laboratorians and the clinical care team to be vigilant about mechanically-induced hemolysis from the extracorporeal circuit during hemodialysis. Communication is critical to identify the cause of hemolysis in these patients and prevent unnecessary delays in result reporting. [ABSTRACT FROM AUTHOR]