Design and performance of a dedicated coherent X-ray scanning diffraction instrument at beamline NanoMAX of MAX IV.
Document Type
Article
Source
Journal of Synchrotron Radiation. May2022, Vol. 29 Issue 3, p876-887. 12p.
Subject
*X-ray diffraction
*DEGREES of freedom
*FOCUS (Optics)
*OPTICAL diffraction
*HARD X-rays
*X-ray imaging
Language
ISSN
0909-0495
Abstract
The diffraction endstation of the NanoMAX beamline is designed to provide high-flux coherent X-ray nano-beams for experiments requiring many degrees of freedom for sample and detector. The endstation is equipped with highefficiency Kirkpatrick-Baez mirror focusing optics and a two-circle goniometer supporting a positioning and scanning device, designed to carry a compact sample environment. A robot is used as a detector arm. The endstation, in continued development, has been in user operation since summer 2017. [ABSTRACT FROM AUTHOR]