Density‐weighted concentric rings k‐space trajectory for 1H magnetic resonance spectroscopic imaging at 7 T
Document Type
article
Source
NMR in Biomedicine. 31(1)
Subject
Biomedical Imaging
Bioengineering
Rare Diseases
Adult
Algorithms
Computer Simulation
Female
Humans
Magnetic Resonance Imaging
Male
Metabolome
Phantoms
Imaging
Proton Magnetic Resonance Spectroscopy
echo-planar
spectroscopic imaging
concentric rings
ultra-high field
Medicinal and Biomolecular Chemistry
Biomedical Engineering
Clinical Sciences
Nuclear Medicine & Medical Imaging
Language
Abstract
It has been shown that density-weighted (DW) k-space sampling with spiral and conventional phase encoding trajectories reduces spatial side lobes in magnetic resonance spectroscopic imaging (MRSI). In this study, we propose a new concentric ring trajectory (CRT) for DW-MRSI that samples k-space with a density that is proportional to a spatial, isotropic Hanning window. The properties of two different DW-CRTs were compared against a radially equidistant (RE) CRT and an echo-planar spectroscopic imaging (EPSI) trajectory in simulations, phantoms and in vivo experiments. These experiments, conducted at 7 T with a fixed nominal voxel size and matched acquisition times, revealed that the two DW-CRT designs improved the shape of the spatial response function by suppressing side lobes, also resulting in improved signal-to-noise ratio (SNR). High-quality spectra were acquired for all trajectories from a specific region of interest in the motor cortex with an in-plane resolution of 7.5 × 7.5 mm2 in 8 min 3 s. Due to hardware limitations, high-spatial-resolution spectra with an in-plane resolution of 5 × 5 mm2 and an acquisition time of 12 min 48 s were acquired only for the RE and one of the DW-CRT trajectories and not for EPSI. For all phantom and in vivo experiments, DW-CRTs resulted in the highest SNR. The achieved in vivo spectral quality of the DW-CRT method allowed for reliable metabolic mapping of eight metabolites including N-acetylaspartylglutamate, γ-aminobutyric acid and glutathione with Cramér-Rao lower bounds below 50%, using an LCModel analysis. Finally, high-quality metabolic mapping of a whole brain slice using DW-CRT was achieved with a high in-plane resolution of 5 × 5 mm2 in a healthy subject. These findings demonstrate that our DW-CRT MRSI technique can perform robustly on MRI systems and within a clinically feasible acquisition time.