학술논문

400 MHz/89 mm actively shielded high resolution NMR magnet
Document Type
Periodical
Source
IEEE Transactions on Magnetics IEEE Trans. Magn. Magnetics, IEEE Transactions on. 32(4):2647-2650 Jul, 1996
Subject
Fields, Waves and Electromagnetics
Nuclear magnetic resonance
Magnetic shielding
Superconducting magnets
Coils
Magnetic fields
Superconducting filaments and wires
Temperature
Flanges
Instruments
Boring
Language
ISSN
0018-9464
1941-0069
Abstract
The first actively shielded high resolution NMR magnet was developed and successfully demonstrated. The magnet has excellent field homogeneity in a very large volume and lineshape characteristics exceed the specifications of the standard magnets. Shielding performance is also outstanding, with the 5 G line located only 1 m radially and 1.5 m axially from the magnet center. We are reporting the magnet specifications, details of the design and fabrication, magnet training, field mapping and lineshape results.