Automatic Configuration of a Medical Imaging System to Unknown Delays in Synchronous Input Data Channels
Document Type
Conference
Source
2010 IEEE International Symposium on Circuits and Systems (ISCAS) Circuits and Systems (ISCAS), 2010 IEEE International Symposium on. :1185-1188 May, 2010
Subject
Components, Circuits, Devices and Systems
Biomedical imaging
Delay systems
Data acquisition
Digital systems
Synchronization
Communication channels
Field programmable gate arrays
Clocks
Radioactive decay
Mammography
Language
ISSN
0271-4302
2158-1525
Abstract
The purpose of this paper is to present a new method and structure for the automatic configuration of a digital system to unknown delays in synchronous input data channels. The method makes possible to restore synchronism in node-to-node communication. Synchronism may be lost due to different delays introduced by the various communication channels. The proposed method allows differences in the channel delays larger than one clock cycle. The method targets distributed digital systems with high number of node-to-node communication channels that must transmit high data volume at high data rates and maintain data time coherence. The method is used in a medical imaging system, namely in the Data Acquisition (DAQ) module of the Data Acquisition Electronics (DAE) component of a PEM (Positron Emission Mammography) system. The system is implemented in a multiple FPGA platform (9 Xilinx™ xc2v4000–4bf957 FPGA). The method and its architectural solution have been validated in a hardware prototype. Test results are presented.