Optical calibration of the SNO+ detector in the water phase with deployed sources
Document Type
article
Source
Journal of Instrumentation. 16(10)
Subject
Nuclear and Plasma Physics
Particle and High Energy Physics
Synchrotrons and Accelerators
Physical Sciences
Analysis and statistical methods
Cherenkov detectors
Detector alignment and calibration methods
Neutrino detectors
Engineering
Nuclear & Particles Physics
Physical sciences
Language
Abstract
SNO+ is a large-scale liquid scintillator experiment with the primary goal of searching for neutrinoless double beta decay, and is located approximately 2 km underground in SNOLAB, Sudbury, Canada. The detector acquired data for two years as a pure water Cherenkov detector, starting in May 2017. During this period, the optical properties of the detector were measured in situ using a deployed light diffusing sphere, with the goal of improving the detector model and the energy response systematic uncertainties. The measured parameters included the water attenuation coefficients, effective attenuation coefficients for the acrylic vessel, and the angular response of the photomultiplier tubes and their surrounding light concentrators, all across different wavelengths. The calibrated detector model was validated using a deployed tagged gamma source, which showed a 0.6% variation in energy scale across the primary target volume.