68Ga-PSMA-11 PET/MRI: determining ideal acquisition times to reduce noise and increase image quality
Document Type
article
Source
EJNMMI Physics. 7(1)
Subject
Biomedical and Clinical Sciences
Clinical Sciences
Oncology and Carcinogenesis
Biomedical Imaging
Cancer
Clinical Research
Clinical sciences
Biomedical engineering
Medical and biological physics
Language
Abstract
BackgroundIn this study, we investigate the impact of increased PET acquisition time per bed position on lesion detectability, standard uptake value, and image noise in 68Ga-PSMA-11 PET/MRI scans.MethodsScans of twenty patients were analyzed in this study. Patients were injected with 68Ga-PSMA-11 (mean, 5.50 ± 1.49 mCi) and imaged on a 3.0 T time-of-flight PET/MRI. PET images were retrospectively reconstructed using 0.5, 1, 2, 4, 7, and 10 min of PET data. Lesion detectability was evaluated on a 5-point Likert Scale for each lesion in each reconstruction. Quantitative analysis was performed measuring image noise and lesion uptake.ResultsA total of 55 lesions were identified, and lesion detectability increased from 2.07 ± 1.14 for 0.5 min to 4.93 ± 0.26 for 10 min (p