Emergency Energy Transfer System for Implantable Ventricular Assist Devices / 埋め込み型補助人工心臓のための緊急エネルギ維持機構の開発
Document Type
Journal Article
Source
The Proceedings of Ibaraki District Conference. 2020, :505
Subject
Electric power supply
Emergency energy transfer system
Emergency port through the skin
Implantable ventricular assist devices
Transcutaneous energy transmission system
Language
Japanese
ISSN
2424-2683
Abstract
In order to use a transcutaneous energy transmission system safely, we have devised the emergency energy transfer system which is used when the electric power supply is stopped. Emergency port through the skin consists of two copper meshes arranged coaxially and filler silicones and the electric power is supplied to the embedded devices when the continuity needle of double pipe structure is stuck in the emergency port. We determined the dimensions of the emergency port and performed the continuity test. The emergency port was connected to light emitting diode and continuity was confirmed by turning on the light emitting diode. The continuity needle was stuck in the other three places in the emergency port. As a result of the experiment, we checked the current was flowed once. In the current structure, the continuity needle is easy to come off due to the elasticity of filler silicones. So that we will improve the structure of emergency port to prevent the continuity needle coming off.