Accuracy evaluation of CRP measurement by a POCT device in NICU / NICU内でのPOCT機器によるCRP測定の正確性評価
Document Type
Journal Article
Source
日本周産期・新生児医学会雑誌 / Journal of Japan Society of Perinatal and Neonatal Medicine. 2022, 58(1):58
Subject
C-reactive protein(CRP)
Neonatal Intensive Care Unit(NICU)
accuracy evaluation
point-of-care testing device
point-of-care testing(POCT)
Language
Japanese
ISSN
1348-964X
2435-4996
Abstract
At our NICU, we have newly introduced the Yumizen M100 Banalyst(hereinafter referred to as Yumizen M100)as a point of care testing(POCT)device for CRP measurement. Prior to that, we compared the Latessier M POCT analyzer with the central laboratory analyzer, evaluating and examining the correlation of the CRP value and the error of measurement. From April to August 2021, a total of 76 tests were performed on 27 cases measured by Yumizen M100 and Latessier M, and a total of 57 tests were performed on 23 cases measured by all three analyzers, including the one from the central laboratory. The CRP value measured by Yumizen M100 showed a very strong correlation with the value measured by Lattesier M and the central laboratory respectively(r = 0.977, 0.997). When the central laboratory CRP value was compared with Yumizen M100 and Latessier M, the measurement of error was significantly smaller using the Yumizen M100(0.044 vs. 0.079, p = 0.029). The introduction of POCT analyzers that are quick, simple, highly accurate, and capable of measuring small quantities of sample was considered to be useful in the NICU.