학술논문

Rapid and reliable diagnostic method to detect Zika virus by real-time fluorescence reverse transcription loop-mediated isothermal amplification
Original Article
Document Type
Report
Source
AMB Express. April 18, 2018, Vol. 8 Issue 1
Subject
Identification and classification
Methods
Zika virus -- Identification and classification
Microbiological techniques (Diagnosis) -- Methods
Language
English
Abstract
Author(s): Xu-Guang Guo [sup.1] [sup.2] , Yong-Zhuo Zhou [sup.3] , Qin Li [sup.4] , Wei Wang [sup.5] [sup.6] , Jin-Zhou Wen [sup.7] , Lei Zheng [sup.1] , Qian Wang [sup.1] [...]
To detect Zika virus more rapidly and accurately, we developed a novel method that utilized a real-time fluorescence reverse transcription loop-mediated isothermal amplification (LAMP) technique. The NS5 gene was amplified by a set of six specific primers that recognized six distinct sequences. The amplification process, including 60 min of thermostatic reaction with Bst DNA polymerase following real-time fluorescence reverse transcriptase using genomic Zika virus standard strain (MR766), was conducted through fluorescent signaling. Among the six pairs of primers that we designate here, NS5 was the most efficient with a high sensitivity of up to 3.3 ng/[mu]l and reproducible specificity on eight pathogen samples that were used as negative controls. The real-time fluorescence reverse transcription LAMP detection process can be completed within 35 min. Our study demonstrated that real-time fluorescence reverse transcription LAMP could be highly beneficial and convenient clinical application to detect Zika virus due to its high specificity and stability.