학술논문

Analysis of Position-Sensitive Capacitive Frisch-Grid TlBr Detectors
Document Type
Conference
Source
2021 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2021 IEEE. :1-9 Oct, 2021
Subject
Communication, Networking and Broadcast Technologies
Nuclear Engineering
Signal Processing and Analysis
Temperature measurement
Performance evaluation
Temperature sensors
Temperature dependence
Three-dimensional displays
Detectors
Crystals
TlBr semiconductor
Position-sensitive Capacitive Frisch-Grid
Room-temperature gamma-ray detector
Language
ISSN
2577-0829
Abstract
This work presents results from analyzing position-sensitive capacitive Frisch-grid (PS-CFG) TlBr gamma-ray detectors. As a room-temperature semiconductor detector, TlBr exhibits a high atomic number, high density, and low Fano factor compared to other material. The use of the 3D position sensing technique provides information on the crystal uniformity. This technique presents the spatial variation in the histograms of the anode amplitude versus the cathode-to-anode ratio, necessary for depth correction, for detectors fabricated with 5×5×12 mm 3 TlBr crystals at room temperature operating at continuous bias (1.5 kV) over the course of nine months. This work also presents the temperature dependence of the leakage current for a TlBr PS-CFG detector over the temperature range of -20 to +60 °C, which varies from