학술논문

A graphene-based pH sensor on paper for human plasma and seawater
Document Type
Conference
Source
2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) Engineering in Medicine and Biology Society (EMBC), 2019 41st Annual International Conference of the IEEE. :1563-1566 Jul, 2019
Subject
Bioengineering
Electrodes
Plasmas
Glass
Biomedical monitoring
Monitoring
Sensitivity
Language
ISSN
1558-4615
Abstract
The relevance of pH assessment in clinical analysis, environmental and industrial control, has raised the demand for the development of portable, low cost and easy-to-use monitoring systems. This paper proposes a pH sensor printed on a paper support passivated with a solid-ink coating. The sensor exploits the pH sensitivity of a reduced graphene oxide functionalized with 3-(4-aminophenil)propionic acid. The sensor responded in the pH range [4], [10] and had a sensitivity of 46 mV/pH. Tests on human plasma and seawater proved this pH sensor to have similar performances than those of a commercial pH-meter with an uncertainty of 0.1 and 0.2 pH unit in plasma and seawater, respectively.