학술논문

The BC-704 scintillation screen with light readout by wavelength shifting fibers as a highly efficient neutron detector
Document Type
Conference
Source
2011 IEEE Nuclear Science Symposium Conference Record Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE. :423-426 Oct, 2011
Subject
Nuclear Engineering
Bioengineering
Communication, Networking and Broadcast Technologies
Components, Circuits, Devices and Systems
Computing and Processing
Signal Processing and Analysis
Channel estimation
Neutrons
Logic gates
Detectors
Computational fluid dynamics
neutron detection
detection efficiency
lithium-6 based scintillators
zero-crossing method
Language
ISSN
1082-3654
Abstract
The aim of the work is to investigate the revised concept of a neutron detector, offered by Saint Gobain-the BC-704 scintillation screen with a light readout by Wavelength Shifting (WLS) fibers. This detector was developed as a replacement of helium tubes, due to the lack of 3He on the world market. It is characterized by high sensitivity to gamma-rays, but the events from neutrons and gammas are well separated. Measurements with gamma and neutron sources have been performed and the results have been compared with helium-3 counter. According to American National Standards Institute (ANSI), we obtained n/γ discrimination on the level of 10 −10 , whereas absolute neutron detection efficiency was estimated as approximately two times higher than in the case of the reference 3 He tube without the moderator. The tested detector is a sample prototype of one detector configuration of many possible, and that it can be modified to the neutron efficiency specification required.