학술논문

Free-Breathing and Single-Breath Hold Compressed Sensing Real-Time MRI of Right Ventricular Function in Children with Congenital Heart Disease.
Document Type
Article
Source
Diagnostics (2075-4418). Jul2023, Vol. 13 Issue 14, p2403. 11p.
Subject
*CONGENITAL heart disease
*CARDIAC magnetic resonance imaging
*IMAGE quality analysis
*MAGNETIC resonance imaging
*COMPRESSED sensing
Language
ISSN
2075-4418
Abstract
(1) Purpose: to compare right ventricular (RV) functional parameters in children with surgically repaired congenital heart disease (CHD) using single/double breath hold (BH) and free-breathing (FB) real-time compressed sensing (CS) cine cardiac magnetic resonance (cMRI) with standard retrospective segmented multi breath hold (RMB) cine cMRI. (2) Methods: Twenty patients with CHD underwent BH and FB, as well as RMB cine cMRI, at 3T to obtain a stack of continuous axial images of the RV. Two radiologists independently performed qualitative analysis of the image quality (rated on a 5-point scale; 1 = non-diagnostic to 5 = excellent) and quantitative analysis of the RV volume measurements. (3) Results: The best image quality was provided by RMB (4.5; range 2–5) compared to BH (3.9; range 3–5; p = 0.04) and FB (3.6; range 3–5; p < 0.01). The RV functional parameters were comparable among BH, FB, and RMB with a difference of less than 5%. The scan times for BH (44 ± 38 s, p < 0.01) and FB (24 ± 7 s, p < 0.01) were significantly reduced compared to for RMB (261 ± 68 s). (4) Conclusions: CS-FB and CS-BH real-time cine cMRI in children with CHD provides diagnostic image quality with excellent accuracy for measuring RV function with a significantly reduced scan time compared to RMB. [ABSTRACT FROM AUTHOR]