학술논문

Percutaneous removal of a fully expanded SMART stent from the pulmonary artery using various adjunctive techniques.
Document Type
Academic Journal
Source
Japanese Journal of Radiology (JAPAN J RADIOL), Oct2012; 30(8): 688-692. (5p)
Subject
Language
English
ISSN
1867-1071
Abstract
A 36-year-old man with an implanted arteriovenous shunt for hemodialysis was referred for shunt malfunction. Venography of the right upper extremity showed occlusion of the subclavian vein, and a SMART stent was deployed. The fully expanded stent immediately migrated centrally into the left pulmonary artery. As initial efforts to pass a snare over the stent failed, we intentionally passed a microguidewire through stent interstices, snared the end of the microguidewire to create a loop, and pulled the stent/microguidewire/snare combination back into the right ventricle where it separated from the loop because of stent mesh destruction. As the stent remained in the right ventricle, we advanced a 0.035-in. guidewire into the stent lumen, passed an angioplasty balloon over the guidewire, inflated the balloon in the stent, and performed pull-back into the right distal external iliac artery. The stent was then surgically removed via a right inguinal incision without eliciting any complications. Although retrieval of the stent malpositioned in the pulmonary artery was difficult, we retrieved it safely by applying various adjunctive techniques.