Calculation of Minimum-Detectable-Concentration Levels of Radioxenon Isotopes Using the PNNL ARSA System
Document Type
Technical Report
Author
Source
Subject
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
98 NUCLEAR DISARMAMENT, SAFEGUARDS, AND PHYSICAL PROTECTION BETA DETECTION
GAMMA DETECTION
RADIATION MONITORS
XENON ISOTOPES
CTBT
MONITORING
SAMPLING
PERFORMANCE MDC Calculations
Xenon Fission Products
MDC Calculations
Language
English
Abstract
Measurement of xenon fission product isotopes is a key element in the global network being established to monitor the Comprehensive Nuclear-Test-Ban Treaty. The automated Radio-xenon Analyzer/Sampler (ARSA), built by Pacific Northwest National Laboratory, can detect 131mXe, 133mXe, 133Xe, and 135Xe via a beta-gamma counting system. Due to the variable background and sources of these four radio-xenon isotopes, it is important to have as sensitive a detection system as possible and to quantify the Minimum-Detectable-Concentrations (MDC) that such a system will be able to detect to preclude false negative and false positive results. From data obtained from IAR in Germany MDC values for 133Xe were well below the 1 mBq/SCMA as required by the PTS for the Comprehensive Test BAn Treaty [WGB TL-11,1999].