학술논문

Use of a Continuous Wave Laser and Pockels Cell for Sensitive High-Resolution Collinear Resonance Ionization Spectroscopy.
Document Type
Article
Source
Physical Review Letters. 9/25/2015, Vol. 115 Issue 13, p132501-1-132501-6. 6p.
Subject
*CONTINUOUS wave lasers
*POCKELS effect
*HIGH resolution imaging
*COLLINEAR reactions
*RESONANCE ionization spectroscopy
Language
ISSN
0031-9007
Abstract
New technical developments have led to a 2 orders of magnitude improvement of the resolution of the collinear resonance ionization spectroscopy (CRIS) experiment at ISOLDE, CERN, without sacrificing the high efficiency of the CRIS technique. Experimental linewidths of 20(1) MHz were obtained on radioactive beams of francium, allowing us for the first time to determine the electric quadrupole moment of the short lived [t1/2 = 22.0(5) ms] 219Fr Qs = -1.21(2) eb, which would not have been possible without the advantages offered by the new method. This method relies on a continuous-wave laser and an external Pockels cell to produce narrow-band light pulses, required to reach the high resolution in two-step resonance ionization. Exotic nuclei produced at rates of a few hundred ions/s can now be studied with high resolution, allowing detailed studies of the anchor points for nuclear theories. [ABSTRACT FROM AUTHOR]